DIOP AIR

Description

Preloaded Hydrophobic intraocular lens

– Haptic angulation : 6°
– Optic size : 6.25 mm
– Haptic design : One piece lens 360° square Hydrophobic
– Material : Hydrophobic acrylic
– Available in size : 12 – 13 – 14 mm
– Sterile Under Blister
– Preloaded inside in own injector

MORE SECURE, NO MANIPULATION, THE BEST COMPROMISE FOR YOUR DIABETIC CATARACT

Coating procedure

① : DIOP Air is located into the cartridge, protected from any ejection by a small plastic sheet which folds in between the two wings of the cartridge at their closure. 
You can notice two holes in it
 
② :   use 1% / 1,2% HaNa or   2% HPMC MethylCellulose viscoelastic  and  inject  one nut   by the first hole
 
③ :  idem by the 2nd hole in that two first cases, the viscoelastic remains above DIOP Air
 
④ : The most important thing now is to inject some viscoelastic underneath DIOP Air coming by the right side of the blue plunger cushion with the visco cannula.

(If you forget to do that DIOP Air will remain stuck inside the cartridge as hydrophobic material doesn’t really like the plastic)

 
⑤ : one viscoelastic nut inside the cartridge barrel
 
Notice:  No more viscoelastic volume is necessary versus a standard IOL.